📈III. Impedance Check
Required PPE: Gloves, Lab Coat
Check Impedance
Select the omega (Ω) symbol in the bottom left corner

Locate the display switch under the researcher screen and press the "Select" button on the right side of the device to switch the participant's computer display to the Mac's
Option 1: Participant PC
Option 2: EEG Mac
Identify and adjust the high-impedance electrodes (red and yellow)
All electrodes ideally should be either green or blue
Gently move the electrode around to make better contact the scalp
If necessary, add one small drop of electrolyte solution using a pipet onto the sponge of the electrode
Do not overuse electrolyte solution or soak the participant's scalp to avoid bridging electrodes
Select "Save" in the bottom right when finished
Switch the participant monitor back to its PC screen